5.2 Michelson interferometer
We came across the Michelson interferometer in Section 2.4 when we discussed the phase relation at a
beam splitter. The typical optical layout of the Michelson interferometer is shown again in Figure 28: a
laser beam is split by a beam splitter and send along two perpendicular interferometer arms. The four
directions seen from the beam splitter are called North, East, West and South. The ends of these arms
(North and East) are marked by highly reflective end mirrors, which reflect the beams back into
themselves so that they can be recombined by the beam splitter. Generally, the Michelson
interferometer has two outputs, namely the so far unused beam splitter port (South) and the input port
(West). Both output ports can be used to obtain interferometer signals, however, most setups
are designed such that the signals with high signal-to-noise ratios are detected in the South
port.
The Michelson interferometer output is determined by the laser wavelength
, the reflectivity and
transmittance of the beam splitter and the end mirrors, and the length of the interferometer arms. In many
cases the end mirrors are highly reflective and the beam splitter ideally a 50:50 beam splitter. In that case,
we can compute the output for a monochromatic field as shown in Section 2.4. Using Equation (20) we can
write the field in the South port as
We define the common arm length and the arm-length difference as
which yield
and
. Thus, we can further simplify to get
The photo detector then produces a signal proportional to
This signal is depicted in Figure 29; it shows that the power in the South port changes between zero and
the input power with a period of
. The tuning at which the output power drops to zero is
called the dark fringe. Current interferometric gravitational-wave detectors operate their Michelson
interferometer at or near the dark fringe.
The above seems to indicate that the macroscopic arm-length difference plays no role in the Michelson
output signal. However, this is only correct for a monochromatic laser beam with infinite coherence length.
In real interferometers care must be taken that the arm-length difference is well below the coherence length
of the light source. In gravitational-wave detectors the macroscopic arm-length difference is an important
design feature; it is kept very small in order to reduce coupling of laser noise into the output but needs to
retain a finite size to allow the transfer of phase modulation sidebands from the input to the
output port; this is illustrated in the Finesse example below and will be covered in detail in
Section 6.4.